The Need
Today's demanding trace impurity detection applications require the utmost in accuracy from the gas purity analyzers monitoring them. To ensure that these high technology analyzers are operating optimally, it is often necessary, or simply desired, to periodically verify their calibration.
The Solution
The Trace Analytical MGB1000 Micro Gas Blender generates consistent and reliable fixed-point gas concentrations at blend ratios up to 4,545 to 1. Applications include calibration and/or certification of gas analysis instrumentation.
The MGB1000 is ideal for calibrating ta3000 and ta5000 gas analyzers with ppm or ppb mixes of a single component or mixed component gas standard. The micro volume flow path enables single step blends quickly and reliably, without consuming large quantities of zero gas.
The MGB1000 is specially engineered for long term stability. All tubing and interconnecting fittings in the primary flow path of the MBG1000 are constructed from 316 stainless steel. Electronic pressure gauges are stable to within 0.1% of full-scale reading per year. As an extra precaution, each gas inlet line is protected from particulate matter by a 2 micrometer particle filter. The pressure-to-flow relationship of the capillary restrictors does not change with time, thereby providing stability virtually over the lifetime of the product.
Advantages
Front Panel: | CAL-gas pressure adjustments and display Blended gas pressure adjustment and display Zero gas pressure display Calibration gas restrictor compartment |
Back-Pressure Regulators | |
CAL-gas pressure: | 100 psig (6.9 bar) max. |
Blend gas pressure: | 15 psig (1 bar) max. |
Pressure Gauges | |
Maximum Operating Pressure: | 100 psig (6.9 bar) |
Linearity: | + .025 psig (0.0017 bar) |
Temperature Effect: | (22-28°C) + 0.5% of reading |
Repeatability: | + 0.05 psig (0.0035 bar) |
1 year stability: | + 0.1 psig (0.0069 bar) |
Readout: | + 0.1 psig (0.0069 bar) |
Construction, diaphragm and Housing: | 316 SS |
Flow Rate Reproducibility Over 1 Year | |
Flow Rate < 0.1 mL/min: | + 2% of reading |
Flow Rate > 0.1 mL/min: | + 1 % of reading |
Electrical | |
Standard: | 110 VAC, 50-60 Hz |
Optional: | 230 VAC, 50-60 Hz |
Mechanical | |
Internal piping, primary flow path: | 316 stainless steel |
Internal volume, primary flow path: | < 0.5 mL |
Rear Panel gas ports: | 1/16" compression |
Dimensions: | 5.3" H x 16.8" W x 14.0" D (133 mm x 427 mm x 356 mm) |
Weight | |
Net: | 15.5 lb (6.8 kg) |
Shipping: | 18.0 lb (8.2 kg) |
Application Note
• Use and Accuracy
Copyright © 2020 Labindia-Instruments | SEO and Web Design by Product Search India
ProMaxion Process Mass Spectrometer for the PAT (Process Analytical Technology) Initiative | ta3000 Gas Analyzers
ta5000 Gas Analyzers | ta7000 Gas Purity Monitors | MGB1000 Micro Gas Blender | Sigma 4000 Multipoint Stream Selector