Test Method
Determines oxidation stability of mineral transformer oils by measuring
the amount of sludge and acid formed under prescribed accelerated
aging conditions.
Oxidation Stability Bath
. Conforms to ASTM D2440 specifications
. Microprocessor temperature control with digital display and
over temperature cut-off
. Six-sample testing capacity
Constant temperature oil bath for testing oxidation stability of mineral
insulating oils. Immerses six oil receptacles at the required depth per
ASTM specifications at 110°C ± 0.5°C, and controls oxygen flow to each
sample at the rate of 1L/h ± 0.1L/h through six independent flow-meters
mounted on a common manifold. Insulated double-wall stainless steel
bath has microprocessor temperature control with °C/°F switchable
digital setpoint and display. Operator and equipment are protected by an
over temperature control circuit which automatically interrupts power to
the unit when bath temperature exceeds a programmed cut-off point.
Communications software (RS232, etc.), ramp-to-set and other enhanced
features are available as extra cost options. Contact your Koehler
representative for information. Order bath thermometer drying tower and
catalyst separately
Specifications
Conforms to the specifications of : ASTM D2440
Capacity: Six samples
Temperature Range: ambient to 260°F (127°C)
Circulator: 1/20hp impeller
Bath Capacity/Medium: 2.5 gal (9.5L) white technical oil
Electrical Requirements: 115V 60Hz, Single Phase, 8.1A ;
220-240V 50/60Hz, Single Phase, 4.2A
Dimensions
lxwxh,in.(cm) : 14x15x22 (36x38x56)
Net Weight: 31 lbs (14.1kg)
Included Accessories
Oil Receptacle and Head (6)
Shipping Information
Shipping Weight: 61 lbs (27.7kg)
Dimensions: 14.4 Cu. ft.